LAMINAR AIR FLOW

Concept:

Laminar air flow is referred to as a system that provides continuous air flow that is uniform in both direction and velocity. Laminar air flow cabinets facilitate unidirectional air flow in a confined space. They create the particulate-free working area by emitting air through a filtration system and exhausting it across the work surface.

They are enclosures designed to flow HEPA-filtered air with uniform velocity and direction. Due to this unidirectional airflow, a sterile environment is maintained while preventing cross-contamination.

Vertical laminar airflow: Being the more common type, here the air flows from top to bottom. The blower and the filters are staged on top of the device. From the enclosure’s front, the air flows out after hitting the work surface.

TECHNICAL SPECIFICATIONS :
Name of the Equipment Thermotek Laminar Air Flow
Work Table Stainless Steel (304)
Construction STD: working surface stainless steel, other construction mild steel powder coated. Side wall transparent acrylic
GMP: working surface & other construction made by stainless steel. Side wall transparent acrylic
Stand Separate fix stand. Max height 650 mm
Filter Type HEPA filter
Efficiency 99.99%
Particle Retention 0.3 Micron
Cleanliness Class 100
Air Flow Velocity 0.3 – 0.5 m/s
Illumination of Work Surface Fluorescent tube light – High intensity, low wattage >800 lux
Blower-Motor Assembly Dynamically balanced
Door Foldable acrylic
Noise Level 65 decibels
Static Pressure Measurement Magnehelic Gauge (0 to 25 MMW)
Pre Filter Disposable, Non-washable polyester fiber
Ultra Violet Tube Light 1, 2, 3 feet, 8–30 Watt – PHILIPS – HOLLAND make – 1 No. top mount
Power Works on 230 V AC single phase 50Hz.
ORDERING INFORMATION FOR THERMOTEK LAMINAR AIR FLOW:
Cat. No. (STD Model) Cat. No. (GMP Model) Chamber Size (WxDxH in ft) Air Flow
LAF-MS-0002 LAF-SS-0002 2x2x2 ft Vertical
LAF-MS-0003 LAF-SS-0003 3x2x2 ft Vertical
LAF-MS-0004 LAF-SS-0004 4x2x2 ft Vertical